• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR® NEXT EZ; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR® NEXT EZ; BLOOD GLUCOSE METER Back to Search Results
Model Number 9628
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.The contour® next ez meter does not have an expiration date.Therefore, no information was captured in section d4 (expiration date).The warranty period of the meter is 5 years from the date of the original purchase.The contour® next ez meter with sku # 9628 and serial # 6966458 was manufactured before the implementation of udi #.Therefore, only di # has been captured in section d4.The kit # associated with the meter is not available.If the information is available at the later date, it will be provided in the supplemental report.
 
Event or Problem Description
The customer reported that her blood glucose readings were not being stored in the memory of the contour® next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the contour® next ez meter with serial # (b)(6) for evaluation.No test strips were returned.Therefore, the returned meter and the in-house contour® next test strips from lot # 3mheg02a were used for in-house testing, using blood spiked with glucose at 135 mg/dl, as determined by the ysi instrument in five replicates.All tests recovered within the fit-for-use limits.The results obtained with the in-house testing were properly stored in the meter's memory.
 
Additional Manufacturer Narrative
Sections d4 (kit lot #) and g4 (510k#) have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR® NEXT EZ
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PHC CORPORATION (REGISTRATION #: 3004167884)
110 oaza-inoshiri-aza-
nishiueno
wakimachi, mima tokusima 779-3 603
JA   779-3603
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key21146570
Report Number1810909-2025-00008
Device Sequence Number15785337
Product Code NBW
UDI-Device Identifier00301939628014
UDI-Public00301939628014
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K191286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9628
Device Catalogue Number83785544
Device Lot NumberDW01H015P
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2024
Supplement Date Received by Manufacturer02/12/2025
04/02/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date03/13/2025
04/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age85 YR
Patient SexFemale
-
-