| Model Number |
9628 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.The contour® next ez meter does not have an expiration date.Therefore, no information was captured in section d4 (expiration date).The warranty period of the meter is 5 years from the date of the original purchase.The contour® next ez meter with sku # 9628 and serial # 6966458 was manufactured before the implementation of udi #.Therefore, only di # has been captured in section d4.The kit # associated with the meter is not available.If the information is available at the later date, it will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reported that her blood glucose readings were not being stored in the memory of the contour® next ez meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the contour® next ez meter with serial # (b)(6) for evaluation.No test strips were returned.Therefore, the returned meter and the in-house contour® next test strips from lot # 3mheg02a were used for in-house testing, using blood spiked with glucose at 135 mg/dl, as determined by the ysi instrument in five replicates.All tests recovered within the fit-for-use limits.The results obtained with the in-house testing were properly stored in the meter's memory.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections d4 (kit lot #) and g4 (510k#) have been updated.
|
| |
|