• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES KONECT RESILIA AORTIC VALVED CONDUIT AVC; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES KONECT RESILIA AORTIC VALVED CONDUIT AVC; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11060A27
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was learned via the patient registry that a patient with a 27mm 11060a aortic valved conduit underwent redo aortic valve replacement after an implant duration of four (4) months due to unknown reasons.The explanted valve was replaced with a 23mm 11060a aortic valved conduit.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
A device history record (dhr) review was not performed, as no information regarding a device failure mode was provided.Furthermore, there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, a definitive root cause cannot be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KONECT RESILIA AORTIC VALVED CONDUIT AVC
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key21146880
Report Number2015691-2025-00313
Device Sequence Number9134962
Product Code LWR
UDI-Device Identifier00690103201390
UDI-Public(01)00690103201390(17)270327
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P150048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number11060A27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer02/12/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date02/18/2025
Date Device Manufactured03/27/2024
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient Age63 YR
Patient SexMale
-
-