MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 LEFT VENTRICULAR ASSIST DEVICE; Ventricular (assist) bypass

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 LEFT VENTRICULAR ASSIST DEVICE; Ventricular (assist) bypass Back to Search Results
Model Number 106523
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Specific patient information and device serial number are documented as unknown.Details are listed in the attached article.Device was implanted at time of event.Date of event has been entered as 01jan2023 since the patients were implanted between 2013 to 2023.Silvia ajello et al., asaio journal70.11: e150-e152.Doi: 10.1097/mat.0000000000002201.Cardiac intensive care unit, istituto di ricovero e cura a carattere scientifico (irccs) san raffaele scientific institute, milan, italy.Manufacturer's investigation conclusion: a direct correlation between heartmate 3 left ventricular assist systems (lvas) and the reported events could not conclusively be established through this evaluation.The serial numbers of the heartmate 3 left ventricular systems in use were not provided.The heartmate 3 left ventricular assist system (lvas) instructions for use (ifu), revision b, is currently available.Section 1, "introduction," lists bleeding as an adverse event which may be associated with the use of heartmate 3 lvas.Section 6, ¿patient care and management¿ (under ¿anticoagulation¿), provides the recommended anticoagulation regimen, including international normalized ratio (inr) range, as well as suggested anticoagulation modifications.No further information was provided.The manufacturer is closing the file on this event.
 
Event or Problem Description
It was reported through the research article titled ¿impella as bridge to durable left ventricular assist device in acute myocardial infarction cardiogenic shock patients?¿ identifying that heartmate 3 (hm3) may be related to bleeding post-left ventricular assist device (lvad) implant that required surgical revision.From 2013 to 2023, 21 patients with cardiogenic shock after acute myocardial infarction were supported with an impella device as a bridge to durable lvad implantation.All patients (100%) were successfully bridged to durable lvad implantation.Seven patients (33%) were implanted with the heartware lvad (hvad; medtronic inc.), and 14 (67%) with the heartmate 3.After lvad implant, 5 patients (24%) required surgical revision for bleeding within 48 hours, and 8 patients (38%) required surgical revision for bleeding within 7 days post-lvad implant.This data first showed that bridging acute myocardial infarction complicated by cardiogenic shock (ami-cs) patients to durable lvad implantation with an impella device was feasible within few weeks from the index event in all patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LEFT VENTRICULAR ASSIST DEVICE
Common Device Name
Ventricular (assist) bypass
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key21148381
Report Number2916596-2025-00230
Device Sequence Number21115902
Product Code DSQ
UDI-Device Identifier00813024013266
UDI-Public(01)00813024013266
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Literature,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number106523
Device Catalogue Number106523
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2025
Initial Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-