MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRELINK; DATA MANAGEMENT SYSTEM Back to Search Results
Model Number 71732-01
Medical Device Problem Codes Unable to Obtain Readings (1516); Application Program Problem (2880)
Health Effect - Clinical Codes Hypoglycemia (1912); Loss of consciousness (2418)
Date of Event 11/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A customer reported an incompatible operating system issue where the low alarm does not sound on the abbott diabetes care (adc) device in use with an honor magic6 lite phone with android operating system version 2.11.2.11107.As a result, the customer experienced a loss of consciousness and was unable to self-treat.The customer required administration of honey by a non-healthcare professional (non-hcp).Additionally, the customer received a glucose injection from a healthcare professional (hcp).There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
An investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer experienced missing low alarm with freestyle librelink application.Per the compatibility guide, use of the honor magic6 lite android operating system, software version 2.11.2.11107 is incompatible with the freestyle librelink application.This guide is available to the user on the websites where the product is launched.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRELINK
Common Device Name
DATA MANAGEMENT SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda 94502-7001
5107495297
MDR Report Key21148490
Report Number2954323-2025-01351
Device Sequence Number9098026
Product Code PZE
UDI-Device Identifier00357599000042
UDI-Public00357599000042
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P160030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number71732-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2025
Initial Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-