Model Number VERADIUS UNITY
Medical Device Problem Code Radiation Output Failure (4027)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported to philips that the wireless footswitch lost connection with the system.The system was not in clinical use at the time of the reported event.There was no reported patient or user harm.Philips has started an investigation of this complaint.
 
Additional Manufacturer Narrative
Philips has investigated this complaint.According to the additional information collected, the system was not in clinical use when the issue was identified.A philips field service engineer (fse) inspected the system onsite and identified that wireless footswitch had lost the connection.Upon troubleshooting fse identified the cause due to weak battery.To resolve the issue, fse reconnect the footswitch.After reconnecting the wireless footswitch, the system was returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude a product malfunction with the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, new information has been received which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.As per the instructions for use, the user of the product must institute a user routine checks program.The user of the product shall make sure that all checks and actions have been satisfactorily completed before using the product for its intended purpose.It is instructed to not use the product for any application until the user is sure that the user routine checks have been satisfactorily completed, therefore, as the device was outside of use at the time the issue was identified, the user would identify this issue prior to use.This complaint is not related to the connection issue, but it was related to the battery problem.Based on the investigation results, philips concludes that the complaint is not reportable.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERADIUS
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN  410501
Manufacturer (Section G)
PHILIPS INDIA LTD.
plot no. b-79, midc,
phase-ii, chakan
pune 41050 1
IN   410501
Manufacturer Contact
marjan vos
veenpluis 6
best 5684 -PC
NL   5684 PC
MDR Report Key21149499
Report Number3003768277-2025-000315
Device Sequence Number9098060
Product Code OWB
UDI-Device Identifier00884838058194
UDI-Public(01)00884838058194
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K142708
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 04/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVERADIUS UNITY
Device Catalogue Number718132
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2025
Supplement Date Received by Manufacturer03/31/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date04/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/17/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1