MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATUNE CRS FB TIB BASE SZ 5 CEM; ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATUNE CRS FB TIB BASE SZ 5 CEM; ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 150640005
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/31/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had a uti roughly one month ago, and the surgeon believes it seeded her total knee as the patient had been doing quite well prior to this event.All well fixed, well aligned and well functioning implants were removed and revised in a one stage revision knee for infection.There was no delay nor patient harm noted.No allegation was made against any component.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Added: d10 concomitant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUNE CRS FB TIB BASE SZ 5 CEM
Common Device Name
ATTUNE STEMMABLE TIBIA IMPLANTS : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21149593
Report Number1818910-2025-00817
Device Sequence Number13093941
Product Code JWH
UDI-Device Identifier10603295042792
UDI-Public(01)10603295042792
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160700
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number150640005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/31/2024
Supplement Date Received by Manufacturer01/16/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CEMENTED STEM 14X50MM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE PS FB INSRT SZ 5 5MM; ATTUNE PS FEM RT SZ 5 NAR CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
-
-