| Model Number |
R26153EA |
| Medical Device Problem Codes |
Break (1069); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Additional information: it was reported during the hytersocopy procedure was prolonged for 30 minutes on (b)(6) 2024, the surgeon was not able to complete the procedure, the patient was referred to gyn.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reflected on adverse event, in section b1, b5 and in section h1.This event is filed under internal complaint id (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It is unknow if the device will be returned to the manufacturing site for investigation.In the event the device returns and relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.Cross reference complaint id: (b)(4).This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported during a medical procedure on (b)(6) 2024, the physican was trying to get theproduct through the cervix and the scope broke.The hysteroscope sheath and hopkins telescope 30°, 2.9 mm, 30 cm were attached (both instruments were being used during the procedure).It was not reported if there was any harm to the patient.No additional information was provided after multiple attempts.Cross reference mdr: 9610617-2025-00026.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in section d9 to reflect that the product was returned on february 19, 2025 for evaluation.The investigation is not completed yet.The investigation results are pending.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per investigation by the manufacturing site: the breakage of the guide tube is a consequence of the bent optics.Since the age is unknown, the shaft may show signs of wear and tear.This can also change the material properties.For example, attempting to bend the shaft back may also cause it to break.Since the main cause can be traced back to the optics, the investigation focuses on the optical data.Cross reference complaint id: (b)(4).This event is filed under internal complaint id (b)(4).
|
| |
|
Event or Problem Description
|
|
Cross reference mdr: 9610617-2025-00026.
|
| |
|