MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3502800 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A female patient who underwent an unspecified breast surgery with a mentor smooth round moderate plus profile 800cc saline prosthesis experienced postoperative right-sided deflation, which was discovered during the surgical procedure.Consequently, the patient underwent bilateral implant replacement with unspecified implants on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on february 27, 2025: the product was returned to mentor for evaluation.Mentor then conducted visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned device revealed that no damage or anomalies were observed on the sm mpps dv 800cc breast implant.Leak testing was performed, in accordance with mentor procedures, a tear and an area of missing material were observed on the anterior view measuring approximately 0.1 cm and 0.1 x 0.1 cm respectively.Microscopic examination was performed on the edges of the tear and missing material, and parallel striations were found in the whole area of the tear and missing material.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|