• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD NEEDLE SFTYGLD 23X1 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD NEEDLE SFTYGLD 23X1 RB; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Catalog Number 305902
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B.3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd needle sftygld 23x1 rb needle was clogged / blocked.Verbatim#: the following information was provided by the initial reporter: rcc received a complaint via phone.Pir attached.25x1 safety glade.Having trouble pushing the drug through the needle.Just had to re-stick the patient¿s 4th different time over the last month (b)(6) lot rejn0115.(b)(6) lot 40954508.Additional information provided: 1.What is the date of event for both material? this has happened a lot over the past several months.2.Please share the correct lot number for material 305916 lot rejn0115 we also have lot 3333815 material 305902 lot 4095458 3.Describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.No patient harm other than having to stick them multiple times.4.Any sample available for investigation? if yes, are you able to provide the address of the facility for us to ship the return label? please mention the material for available sample.(b)(6).5.Please share the captured photo of the event if available.
 
Additional Manufacturer Narrative
One sample received for investigation.Through visual inspection, no defects or issues observed.Functional testing was performed, solution expelled with normal flow.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Based on the quality team's investigation, we are not able to identify a root cause related to our manufacturing process at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEEDLE SFTYGLD 23X1 RB
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key21152558
Report Number1213809-2024-01055
Device Sequence Number13070516
Product Code FMI
UDI-Device Identifier30382903059028
UDI-Public(01)30382903059028(11)240404(17)290331(10)4095458
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K951254
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number305902
Device Lot Number4095458
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2024
Supplement Date Received by Manufacturer02/06/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date02/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-