| Catalog Number |
305902 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B.3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd needle sftygld 23x1 rb needle was clogged / blocked.Verbatim#: the following information was provided by the initial reporter: rcc received a complaint via phone.Pir attached.25x1 safety glade.Having trouble pushing the drug through the needle.Just had to re-stick the patient¿s 4th different time over the last month (b)(6) lot rejn0115.(b)(6) lot 40954508.Additional information provided: 1.What is the date of event for both material? this has happened a lot over the past several months.2.Please share the correct lot number for material 305916 lot rejn0115 we also have lot 3333815 material 305902 lot 4095458 3.Describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.No patient harm other than having to stick them multiple times.4.Any sample available for investigation? if yes, are you able to provide the address of the facility for us to ship the return label? please mention the material for available sample.(b)(6).5.Please share the captured photo of the event if available.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample received for investigation.Through visual inspection, no defects or issues observed.Functional testing was performed, solution expelled with normal flow.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Based on the quality team's investigation, we are not able to identify a root cause related to our manufacturing process at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|