FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; BREATHING CIRCUIT Back to Search Results
Model Number RT266
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported on behalf of a healthcare facility in china that an mr290v vented autofeed humidification chamber provided with an rt266 infant dual-heated evaqua2 breathing circuit was found leaking water prior to patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Fisher & paykel (f&p) healthcare is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject mr290v vented autofeed humidification chamber was not returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.Our investigation is therefore based on the limited information and photographs provided by the healthcare facility and our knowledge of the product.Results: the provided photographs shows small amount of water located on the mr290v humidification chamber base.No visible damage can be determined from the provided photograph.Conclusion: without the return of the subject device, we are unable to confirm the cause of the reported event.Every mr290v chamber is pressure tested following the manufacturing process and any holes, cracks, or leaks would be identified during this process.In addition, the pressure test is followed by a visual inspection of each chamber.Any chamber which fails these tests is rejected.The subject mr290v humidification chamber would have met the required specifications at the time of production.The user instructions for the rt266 breathing circuit state the following: - do not use the chamber if the seals are not intact when received, or if it has been dropped.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- ensure there is a water supply connected to the chamber and that water is present within the chamber.- appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.- ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.- visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in china that an mr290v vented autofeed humidification chamber provided with an rt266 infant dual-heated evaqua2 breathing circuit was found leaking water prior to patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
BREATHING CIRCUIT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key21154849
Report Number9611451-2025-00032
Device Sequence Number15750629
Product Code BZE
UDI-Device Identifier09420012431103
UDI-Public(01)09420012431103(10)2103078395(11)240409(17)290409
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K103767
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number2103078395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2024
Supplement Date Received by Manufacturer02/14/2025
Initial Report FDA Received Date01/14/2025
Supplement Report FDA Received Date03/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-