| Model Number |
RT266 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china that an mr290v vented autofeed humidification chamber provided with an rt266 infant dual-heated evaqua2 breathing circuit was found leaking water prior to patient use.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel (f&p) healthcare is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the subject mr290v vented autofeed humidification chamber was not returned to fisher & paykel (f&p) healthcare in new zealand for evaluation.Our investigation is therefore based on the limited information and photographs provided by the healthcare facility and our knowledge of the product.Results: the provided photographs shows small amount of water located on the mr290v humidification chamber base.No visible damage can be determined from the provided photograph.Conclusion: without the return of the subject device, we are unable to confirm the cause of the reported event.Every mr290v chamber is pressure tested following the manufacturing process and any holes, cracks, or leaks would be identified during this process.In addition, the pressure test is followed by a visual inspection of each chamber.Any chamber which fails these tests is rejected.The subject mr290v humidification chamber would have met the required specifications at the time of production.The user instructions for the rt266 breathing circuit state the following: - do not use the chamber if the seals are not intact when received, or if it has been dropped.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- ensure there is a water supply connected to the chamber and that water is present within the chamber.- appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.- ensure appropriate ventilator or flow source alarms are set before connecting breathing sets to patient.- visually inspect breathing sets for damage (e.G.Crushed tube or cracked connector) before use and replace if damaged.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in china that an mr290v vented autofeed humidification chamber provided with an rt266 infant dual-heated evaqua2 breathing circuit was found leaking water prior to patient use.There was no patient involvement.
|
| |
|