The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received no information alleging.There was no report of medical intervention.No additional information can be requested at this time.The manufacturer was made aware of this complaint through a representative of the customer.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.During the evaluation, no secondary findings was observed.The device's downloaded logs were reviewed by the third-party service center.There were no errors code found.The third-party service center concludes that they couldn't confirm the customer's allegation and there was no visible foam degradation and unit got scrapped.
|