• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTR ANCH,BIO-COMP S-TAK; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTR ANCH,BIO-COMP S-TAK; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number SUTR ANCH,BIO-COMP S-TAK
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a surgery after broström the implant pulled out after inserting the device into the drill channel.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with just one device (in this surgery they normally use two devices).It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH,BIO-COMP S-TAK
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21157048
Report Number1220246-2025-09456
Device Sequence Number15361388
Product Code MAI
UDI-Device Identifier00888867097018
UDI-Public(01)00888867097018(11)240423(17)280331(10)15238547
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K140855
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTR ANCH,BIO-COMP S-TAK
Device Catalogue NumberAR-1934BCF-2-1
Device Lot Number15238547
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer06/03/2025
Initial Report FDA Received Date01/15/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-