| Catalog Number |
A0756225G0S03 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation and the lot number was not provided for retained-product testing and/or dhr review.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a waveone gold small 6-file ster 25mm file broke during use.The broken part was not retrieved and was incorporated into the filling.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.The batch number is unknown, dhr cannot be reviewed.Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique, equipement (motor issue), overuse (number of uses not communicated), excessive wear, or material issue (no analysis of the broken fragments possible).Root causes are not identified.
|
| |
|
Search Alerts/Recalls
|