EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 3.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-010-01-0235 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Scar Tissue (2060); Osteolysis (2377)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: (2221302) 02-012-44-3511 - logic tibia implant psc insert, sz 3.5, 11mm.(2819341) 02-012-41-3535 - logic tibia traptray cem sz 3.5f/3.5t.200-02-35 - three peg patella 35mm.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00718.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 105 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear and aseptic loosening.Revision surgery operative notes were provided.Post operative diagnosis noted periprosthetic osteolysis of internal prosthetic, left knee joint.Intraoperatively the surgeon observed the femoral component to be significantly loose.A lateral defect was noted distally and posteriorly and noted by the surgeon to be consistent with preoperative radiographs.Scar tissue was observed around the tibial component.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: g3/g4, h6; the following sections were corrected: h6.Mdr section codes updated/corrected: b, c, e.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|