| Catalog Number |
NT3211 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported periimplantitis at tooth site #32 and the implant was removed.Inflammation was reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).H6: event problem codes: patient code: 1690.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g6: type of report h1: type of reportable event h2: follow up type h10: additional narrative: zimvie complaint number corrected h3 other text : summary investigation.
|
| |
|
Search Alerts/Recalls
|