Complaint allegation is confirmed.One unpackaged ar-12990n-1, scorpion¿ needle, batch number: 15291754, was received for investigation.The device was alleged to be broken inside scorpion, ar-12990, batch: 14478202.Visual evaluation of the returned scorpion noted a fragment of a foreign material in the suture slot.Functional testing was performed by attempting to insert a known good, ar-12990 knee scorpion needle, into the returned device, the needle was met with resistance, and it was found that the cannulated shaft of the instrument is obstructed.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during the firing of the needle thus, breaking the needle and causing a blockage within the shaft.Dfu-0392-sub-eo warns against the usages listed to help avoid needle breakage and potential patient injury.
|