• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 CANNULATED IMPACTOR HANDLE; SHOULDER INSTRUMENT - IMPACTOR HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 CANNULATED IMPACTOR HANDLE; SHOULDER INSTRUMENT - IMPACTOR HANDLE Back to Search Results
Catalog Number 620101141
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of the large bowtie glenoid reamer broke while reaming the glenoid.Fragments were generated and were retrieved.The large blazer fused onto the cannulated handle.There was no mishandling of the product before or after assembling the two items.There was about a 20 minute time delay to try to remove the blazer from the cannulated handle.There was no patient harm.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received "it was reported that the large bowtie glenoid reamer broke while reaming the glenoid.The large blazer fused onto the cannulated handle.There was a delay in case.No harm to patient¿.The product was not returned to medtech orthopedics ; however, photos were provided for review.See ¿(b)(4) device photo ad 12 jan 2025.¿ the device associated with this report was not returned to medtech orthopedics for evaluation.The photographs attached were reviewed, however in the images provided no observations pertaining to the nature of the reported event of jammed or seized could be identified.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs attached don¿t have any evidence to draw a conclusion about the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED IMPACTOR HANDLE
Common Device Name
SHOULDER INSTRUMENT - IMPACTOR HANDLE
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21158362
Report Number1818910-2025-00899
Device Sequence Number7655935
Product Code HWR
UDI-Device Identifier10603295539339
UDI-Public(01)10603295539339(10)201687
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number620101141
Device Lot Number201687
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/31/2024
Supplement Date Received by Manufacturer01/20/2025
Initial Report FDA Received Date01/15/2025
Supplement Report FDA Received Date01/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age75 YR
Patient SexUnknown
-
-