| Model Number |
ONB5SHF |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic inguinal hernia repair procedure, when attempting to reinsert and remove the 5 millimeter scope with the optical method inserted, the housing section also came off.A new trocar was used to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the seal housing was disengaged from the cannula.The circular seal holder was disengaged and the duckbill seal was stretched out.Functional testing found the device failed an air leak test.It was reported that when attempting to reinsert and remove the 5 mm scope, the housing section also came off.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when contact is made with a surgical device during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadve rtent damage to the seal.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|