• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS KIT: 2-L 15 FR X 19 CM; CATHETER, HEMODIALYSIS, IMPLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS KIT: 2-L 15 FR X 19 CM; CATHETER, HEMODIALYSIS, IMPLA Back to Search Results
Catalog Number CS-15242-SP
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "the doctor found the luer hub/fitting has crack during used on the different patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.The associated emdr report numbers are 9680794-2025-00088.
 
Additional Manufacturer Narrative
(b)(4) the customer returned one connector assembly from a hemodialysis kit for analysis.Signs of use were observed.Visual analysis revealed stress markings on both luer hubs and a crack on the blue, venous luer hub.The appearance of this damage is consistent with overtightening or repeated tightening of the hubs.No other defects or anomalies were observed on connector assembly.Both luer connections were functionally tested per the product instructions for use (ifu).The ifu provided with this kit instructs the user, "flush each catheter lumen with sterile normal saline for injection, to establish patency and prime lumen(s) and flush the irrigation tube." a lab inventory syringe filled with water was used to flush each extension line.Leaking was observed from the crack in the blue venous luer hub.A manual tug test confirmed both luer hub were securely attached to their respective lumens.A device history record review was performed, and no relevant findings were identified.The ifu provided with this kit warns the user, "examine catheter and extension lines before and after each treatment for any signs of damage.Repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." the report of damaged luer hubs was confirmed through complaint investigation of the returned sample.Visual analysis revealed stress markings on both luer hubs and a crack in the blue venous luer hub.This damage is consistent with overtightening or repeated tightening of the hubs.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "the doctor found the luer hub/fitting has crack during used on the different patient".They changed a new one to resolve the issue.No harm for the patient.The patient condition is fine.No medical intervention was required.The associated emdr report numbers are 9680794-2025-00088 and 9680794-2025-00087.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS KIT: 2-L 15 FR X 19 CM
Common Device Name
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21159595
Report Number9680794-2025-00087
Device Sequence Number8594585
Product Code MSD
UDI-Device Identifier00801902096326
UDI-Public(01)00801902096326(11)240305(17)270228(10)33F24A0271(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111117
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCS-15242-SP
Device Lot Number33F24A0271
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2024
Supplement Date Received by Manufacturer01/28/2025
Initial Report FDA Received Date01/15/2025
Supplement Report FDA Received Date02/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-