This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: b5, d8, d9, h2, h3, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: water invasion in the cable and the imaging; corrosion on the electrical contacts; cable damage a root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: water invasion in the cable and the imaging; corrosion on the electrical contacts; cable damage.These phenomena were confirmed during inspection at the repair department.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|