Model Number H7493926712350
Medical Device Problem Codes Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Embolism/Embolus (4438)
Date of Event 12/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon rupture occurred requiring additional intervention.During the procedure, a stent was implanted, and a 3.50mm x 12mm nc emerge balloon catheter was placed into the vessel to perform post-dilatation.However, the balloon ruptured during inflation, causing pieces of the balloon to embolize further into the vessel.Attempts were made to remove the separated balloon pieces using a snare device, but not all pieces could be successfully removed.Ultimately, a piece of the balloon material was left in one of the patient's distal right coronary artery (rca) branches.The procedure was completed using an alternate method.No further information was provided regarding the patient's outcome or any additional interventions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21163530
Report Number2124215-2024-84113
Device Sequence Number7831212
Product Code LOX
UDI-Device Identifier08714729846413
UDI-Public08714729846413
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2025
Device Model NumberH7493926712350
Device Catalogue NumberH7493926712350
Device Lot Number0032592420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2024
Initial Report FDA Received Date01/15/2025
Date Device Manufactured10/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-