| Catalog Number |
EP8702H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a cardiovascular procedure in 2024 and suture was used.During the procedure, cv surgeon encountered prolene 7-0 everpoint suture detached from the swage.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: 1.When was the needle detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Ans: needle detached during the surgery.2.What is the source of information for the queries above (e.G., answered by the physician / provided from knowledge as you were present in the case) ans: information given by the scrub nurse.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional h6 investigation findings: c22 - photo review additional h6 component code: g07002 unable to investigate further additional h3 investigation summary: the product was returned to ethicon for evaluation.One detached needle and one needle suture piece outside the winding former were received for analysis.The product code is ep8702h.Upon visual inspection of the detached needle, the swage condition was as expected.The barrel hole of the needle was examined with magnification, and no suture remnant was noted.In addition, the needle-suture piece was visually inspected, and no anomalies or issues related to pull-off were observed during the evaluation.The suture was examined and the end was noted to be cut probably by a surgical instrument.The other section of the suture was not returned for evaluation.The functional test was performed using instron equipment and the pull force was above the minimum requirements.In addition, a photo was provided and it showed, one empty winding former, one detached needle, and one needle suture inside the plastic bag for product code ep8702 a manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached as to what caused the reported complaint.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|