MAUDE Adverse Event Report: GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE Back to Search Results
Catalog Number 303330
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2024
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2024, grifols distributor (b)(4) (on behalf of customer (b)(6), reported a discrepant hiv result for a donor sample tested with procleix ultrio elite assay master lot: (ml) 708054.The initial donor sample tested on (b)(6) 2024 yielded a nonreactive result with an s/co of 0.08.The sample was hiv serology positive in the diasorin assay.The diasorin results were deleted for unknown reasons.The sample was sent to an offsite clia laboratory for additional testing.On (b)(6) 2024, the sample was found nonreactive in the abbott architect anti-hiv assay.The customer accepted the agreement between procleix and abbott results (nonreactive for hiv) as the valid sample interpretation and no sample remained for investigation.No related quality events found for this lot/batch of product and the product met all release criteria.A previous events search did not indicate a trend for hiv sensitivity issues with ultrio elite ml 708054.The root cause of the disagreement between the ultrio elite and abbott results as compared to the diasorin results could not be determined.False positive results in the diasorin assay due to contamination or assay performance could not be ruled out.Another possible root cause of the nonreactive ultrio elite result is low titer hiv.Intermittent reactive results are expected in low titer samples (at or below the 95% lod of the assay).The ultrio elite assay was determined to be working as designed.Risk assessment: severity: critical.A single blood screening false negative (fn) result can cause multiple infections.The severity of a blood screening false negative is critical.Probability: remote.The probability of a false negative result in ml 708054 is remote based on the previous events search.False negative results do occur but are rare.Overall risk: acceptable.Panther assay specific risk analysis for bloodbank assays, gdss-rd-rsk-000043 rev.2.0 and product safety risk management procedure, gdss-ms-sop-000004 version 13.0 were reviewed to assess risk.New risk to patient safety was not identified.There were no related nonconformance's noted in the manufacturing record or as part of a search of the nonconformance records for this lot number.Product impact: no impact to product was identified.Review of the qc release data and a previous events search in the grifols complaint system indicate that the assay is working as designed.Customer impact: no record showed the donation was used.The impact to the customer was questioning the discordant ue result.Following iso 14971, grifols risk assessment is based on severity of a harm or impact to human health and the probability of occurrence of harm or impact to human health.Based on the severity rating of ·critical· and probability rating of ·remote·, the overall risk posed by a false negative ultrio elite result is determined to be acceptable.No further information is expected, this is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLEIX ULTRIO ELITE ASSAY
Common Device Name
HIV-1/HIV-2/HCV/HBV DEVICE
Manufacturer (Section D)
GRIFOLS DIAGNOSTIC SOLUTIONS INC.
4560 horton street
emeryville CA 94608
Manufacturer Contact
jin shi
10808 willow court
san diego, CA 92127
8582020806
MDR Report Key21164075
Report Number2032600-2025-00001
Device Sequence Number4538830
Product Code QHO
UDI-Device Identifier00859882007672
UDI-Public00859882007672
Combination Product (Y/N)N
Initial Reporter CountryID
PMA/510(K) Number
BL125652
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/15/2026
Device Catalogue Number303330
Device Lot Number708054
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2024
Initial Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-