STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5530-G-411 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
12/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding wear involving a triathlon insert was reported.The event for wear was confirmed via evaluation of the provided photographs.Method & results: product evaluation and results: the device was not returned; however, photographs were provided for inspection.The photographs show a recently explanted insert with damage present on the anterior portion of the condyle consistent with delamination and material loss.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability.The device was not returned; however, photographs were provided for inspection.The photographs show a recently explanted insert with damage present on the anterior portion of the condyle consistent with delamination and material loss.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right knee was revised due to instability.A 4x11 cr insert was revised to a 4x11 cs insert.Rep confirmed that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|