ZIMMER BIOMET, INC. BIOMET FINNED PRI STEM 80X15MM; PROSTHESIS, KNEE Back to Search Results
Catalog Number 141320
Medical Device Problem Codes Material Erosion (1214); Fracture (1260)
Health Effect - Clinical Codes Adhesion(s) (1695); Failure of Implant (1924); Synovitis (2094); Metal Related Pathology (4530); Bursitis (4832)
Date of Event 12/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: medical product: biomet ilok pri tib tray 83mm: catalog#141216, lot#ni; unknown femoral component: catalog#ni, lot#ni; unknown articular surface: catalog#ni, lot#ni.G2: foreign: germany.Diligence is in process to determine whether the product is available for evaluation.The investigation is in progress.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an initial right total knee arthroplasty on an unknown date.Subsequently, on an unknown timeframe post-implantation, the patient underwent revision surgery as the connection between the tibial tray and tibial stem fractured.Due diligence is in progress for this event; to date no additional information has been reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: medical product: biomet ilok pri tib tray 83mm: item#41216, lot#456060; unknown 16x79/83mm articular surface: catalog#ni, lot#887550.The investigation is in progress.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: a4; b4; b5; b7; d6a; d9; d10; g3; h2; h6; h11.D10: additional concomitant medical products: unknown femoral component size 75; unknown optipac cement; unknown bone filler the investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient underwent an initial right total knee arthroplasty.Approximately three (3) years post-implantation, the patient underwent revision surgery as the connection between the tibial tray and tibial stem and screw fractured.During the revision surgery metallosis was noted due to the fractured and worn components.The femoral component was well fixed and remains implanted.Due diligence is complete and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h11.H6: proposed component code: mechanical (g04) - stem.Visual examination of the returned product identified wear on the body of the stem implant, with a portion of the fracture tibia still assembled with the stem.The stem screw was found fractured below the head.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.The screw was submitted for further analysis.Analysis determined fracture surface showed relatively smooth surface with indications of ratchet marks and hinge artifacts suggesting a possible fatigue mode of failure.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: revision notes: gray colored fluid encountered & metal debris; femoral component not loose - retained; tibia plate removed noted to be fractured and removed easily; screw head broke off smoothly and was able to grasp the protruding (0.5mm) remnant and unscrew; cone is broken and worn by friction; took approximately half hour to remove stem from cemented shaft.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: right knee arthroplasty exhibited progressive loosening and eventual disassembly of the modular tibial tray and tibial stem components with migration of the tibial tray component.X-rays demonstrate metal-on-metal corrosion of the tibial tray modular connector component at connection with the tibial stem component, metal synovitis adverse local tissue reaction (altr).Root cause was unable to be determined.Reported event was confirmed by review of provided medical records and evaluation of the returned device.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET FINNED PRI STEM 80X15MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21164545
Report Number0001825034-2025-00156
Device Sequence Number9546180
Product Code JWH
UDI-Device Identifier00880304000964
UDI-Public(01)00880304000964(17)300729(10)222960
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K142933
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number141320
Device Lot Number222960
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2024
Supplement Date Received by Manufacturer01/28/2025
05/15/2025
06/27/2025
Initial Report FDA Received Date01/15/2025
Supplement Report FDA Received Date02/25/2025
06/13/2025
06/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age56 YR
Patient SexMale
Patient Weight122 KG
-
-