MEDTRONIC XOMED INC. NIM® EMG - ENDOTRACHEAL TUBE; STIMULATOR, NERVE Back to Search Results
Model Number 8229307
Medical Device Problem Code Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis found that visually, there was no damage in the construction of the device.Functionally, a syringe was used to inject 20cc of air into the cuff balloon and there were no issues during inflation or deflation.The cuff was re-inflated and deflated multiple times and no issues were observed.For further analysis, a leak test was performed by submerging the entire device in water and no bubbles (leakage) were observed.No issues were observed while manipulating the cuff or pilot balloons, and the pilot balloon inflated as intended.A review of the global complaint data showed one similar complaint about this lot number.In the returned condition, there was no out of specification condition that was related to the complaint.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the package was opened for pre-inflation and it was found that the cuff could not be inflated and could not be used.Procedure was completed with another product.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM® EMG - ENDOTRACHEAL TUBE
Common Device Name
STIMULATOR, NERVE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
justin ellis
6743 southpoint drive north
jacksonville, FL 32216
7635265677
MDR Report Key21165936
Report Number1045254-2025-00106
Device Sequence Number11830429
Product Code ETN
UDI-Device Identifier00763000745820
UDI-Public00763000745820
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K925640
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8229307
Device Catalogue Number8229307
Device Lot Number0227468553
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2024
Initial Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-