H3: product analysis found that visually, there was no damage in the construction of the device.Functionally, a syringe was used to inject 20cc of air into the cuff balloon and there were no issues during inflation or deflation.The cuff was re-inflated and deflated multiple times and no issues were observed.For further analysis, a leak test was performed by submerging the entire device in water and no bubbles (leakage) were observed.No issues were observed while manipulating the cuff or pilot balloons, and the pilot balloon inflated as intended.A review of the global complaint data showed one similar complaint about this lot number.In the returned condition, there was no out of specification condition that was related to the complaint.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|