| Catalog Number |
UNK LIFESTENT 5F STENT SYSTEM |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the part of the delivery system allegedly broke off inside the patient's body.Reportedly, the broken piece was retrieved by using a snare device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: as the lot number for the device was not provided, a review of the device history records could not be performed.Investigation summary: neither sample nor image provided which leads to inconclusive result.Based on the information available, the investigation is closed with inconclusive result.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use (ifu) states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the ifu states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the ifu further state: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter (.)'.Regarding insertion and removal difficulty of the delivery system the ifu states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.H10: g3.H11: d4 (medical device catalog #, medical device lot #), h6 (patient, method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the part of the delivery system allegedly broke off inside the patient's body.Reportedly, the broken piece was retrieved by using a snare device.There was no reported patient injury.
|
| |
|