EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-35-4011 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 168 months after a left total knee replacement procedure, the patient has experienced prosthesis wear.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: (b)(6), 02-010-01-0240 - logic femoral ps cem left sz 4.(b)(6), 02-012-41-4030 - logic tibia traptray cem sz 4f/3t.(b)(6), 200-02-32 - three peg patella 32mm.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/ technique.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The prosthesis wear alleged could have been the result of over 13 years of use, however, this cannot be confirmed as the devices have not been revised.The reason for the legal filing may have been due to the inclusion of the polyethylene in the packaging recall.Potential contributions of user or patient-related considerations could not be assessed as the devices remain implanted and no images, radiographs, or relevant clinical information was provided.H6: corrected investigation clinical codes.
|
| |
|