• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC 10 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC 10 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR Back to Search Results
Model Number G31919
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Band deployment difficulty can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." band deployment difficulty can also occur if the trigger cord is not properly seated in the handle assembly.The instructions for use advise the user: "note: knot must be seated into hole or handle will not function properly." a precaution in the instructions for use state: "it is vital that the integrity of the working channel is intact as grooves or other obstructions in the working channel can potentially cause the string to catch, resulting in band deployment difficulty." it is possible that this could lead to band deployment difficulty and damage to the trigger cord as well.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Event or Problem Description
During an endoscopic hemorrhoid ligation procedure in the rectum, the physician used a cook 10 shooter saeed multi-band ligator [mbl].It was reported [that] user successfully deployed five bands, but sixth band cannot be deployed.User changed the endoscope position from retroflexed to forward-viewing position, but the sixth band still cannot be deployed.User retracted the mbl device and installed again and still cannot deploy the sixth band.User changed to a new mbl device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
During an endoscopic hemorrhoid ligation procedure in the rectum, the physician used a cook 10 shooter saeed multi-band ligator [mbl].It was reported [that] user successfully deployed five bands, but sixth band cannot be deployed.User changed the endoscope position from retroflexed to forward-viewing position, but the sixth band still cannot be deployed.User retracted the mbl device and installed again and still cannot deploy the sixth band.User changed to a new mbl device to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in a white plastic bag with an open box/tray from the lot number provided in the report.The label matches the product returned.The irrigation adapter and some bands did not return.Our laboratory evaluation of the returned device could not confirm the report.The device was returned with two bands remaining on the barrel.A functional test was performed on the 2 remaining bands on the barrel.The multi-band ligator device was attached to an endoscope that was placed in a simulated gastrointestinal position with vacuum attached.The endoscope has an accessory channel that is 2.8 mm in diameter (olympus 2.8 gif q20 endoscope).The multi-band ligator barrel was attached onto the end of the endoscope.Simulated varices were placed at the end of the barrel and vacuum pulled and the remaining bands were fired.It was observed that the 2 remaining bands deployed as intended with no difficult deployment observed, but did not constrict as expected.This is due to the functional testing being conducted at room temperature instead of at body temperature as the test was only intended to test for successful deployment.A visual examination of the device was performed.The trigger cord was examined and all twenty deployment beads were present.The beads were verified for correct location using bead inspection plate.The beads were examined using magnification and found to be correctly filled and had no evidence of excess flash.The trigger cord was intact and not broken.The length of the trigger cord was measured within tolerance.An evaluation of the handle wheel movement was performed and the wheel functioned as intended.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device could not confirm the report of difficult deployment.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Band deployment difficulty can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." band deployment difficulty can also occur if the trigger cord is not properly seated in the handle assembly.The instructions for use advise the user: "note: knot must be seated into hole or handle will not function properly." a precaution in the instructions for use state: "it is vital that the integrity of the working channel is intact as grooves or other obstructions in the working channel can potentially cause the string to catch, resulting in band deployment difficulty." it is possible that this could lead to band deployment difficulty and damage to the trigger cord as well.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
MND, LIGATOR
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key21168416
Report Number1037905-2025-00056
Device Sequence Number9356622
Product Code MND
UDI-Device Identifier00827002319196
UDI-Public(01)00827002319196(17)250625(10)W4848675
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2025
Device Model NumberG31919
Device Catalogue NumberMBL-U-10
Device Lot NumberW4848675
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2024
Supplement Date Received by Manufacturer01/20/2025
Initial Report FDA Received Date01/16/2025
Supplement Report FDA Received Date02/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS 290 ENDOSCOPE.
Patient Age53 YR
Patient SexFemale
Patient Weight65 KG
-
-