Catalog Number J594G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H6.Component code: g07002 - device not returned attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: it was confirmed that no needle pieces fell out for the product in which needle breakage occurred.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Were there any patient consequences? 2.Please provide lot number this report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and a suture was used.One needle-sutures was broken, and the suture and needle came off in three other needle-sutures.It was not a control release needle.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Corrected information: g 4.Combination product ¿ this complaint is not related to a combination product.This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 2/24/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, the following was obtained: were there any patient consequences? no.Please provide lot number: unk.Please provide the source or name and title of the external person providing answers to follow-up (external person submitting answers to sales rep).(b)(6).Is the sample available to be returned for evaluation? if so, provide any available tracking information.We regularly contact with sale rep about the device returning.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 5/15/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: g1.Additional information was received: lot number of (b)(4): unk to 100q3d, 100s6c.Additional information: d4 (product code) - the actual product code associated with this event is not known.The possible product codes are reported as follows: product code: 100q3d and product code: 100s6c.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 6/25/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device batch 100s6c and no non-conformances were identified.D4/h4: mfg.Date -4/30/2024 exp.Date - 3/31/2029.Udi: (b)(4).A manufacturing record evaluation was performed for the finished device batch 100q3d and no non-conformances related were identified.D4/h4: mfg.Date -4/25/2024 exp.Date - 3/31/2029.Udi: (b)(4).H3 investigational summary: the product was returned to ethicon for evaluation.One detached needle and one suture piece outside its packaging were received for analysis.Product code j594g.During the visual assessment of the detached needle, the swage and attachment area were as expected.However, marks that appear to be caused by a surgical instrument were observed on the body needle.The barrel hole of the needle was examined with magnification, and no suture remnant was noted.The suture was examined, and on one extreme, a correct insertion mark was observed; however, the force used to detach the needle from the suture could not be determined.As per the returned conditions of the sample, no functional test was performed.Besides that, body fluids along the strand were noted as well.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD UD 18IN 4-0 S/A PS-5 PRM
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21169583
Report Number2210968-2025-00616
Device Sequence Number4336456
Product Code GAM
UDI-Device Identifier10705031041578
UDI-Public10705031041578
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberJ594G
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2025
Supplement Date Received by Manufacturer01/16/2025
02/05/2025
04/17/2025
05/30/2025
Initial Report FDA Received Date01/16/2025
Supplement Report FDA Received Date01/16/2025
02/24/2025
05/15/2025
06/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-