| Catalog Number |
MP1000 CHINA |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd maxplus needleless connector was occluded.The following information was provided by the initial reporter: the patient was seen at our hospital for 11 hours for chest compression and was given fluids on (b)(6) 2024, as prescribed, and after using an infusion set connected to a needleless connector, the fluids did not pass through, and a replacement needleless connector was given.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: if a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
A mp1000 china product was not available for investigation of (b)(4); however, the customer confirmed that the complaint sample was from lot 23105040.The feedback provided by the customer states that, "an infusion set connected to a needleless connector, the fluids did not pass through".No further information was available to assist the investigation in this instance.The details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 23105040 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.The root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.However, previous investigations have identified that certain designs of male luer can cause flow restriction or occlusion as they can grip onto the piston of the maxplus preventing its proper collapse.This can sometimes be resolved by disconnecting and reattaching the same luer connection which may reposition the luer against the piston and improve the flow, or alternatively by changing the connecting male luer.In this instance, without the complaint sample to examine it has not been possible to conclusively link this feedback to a specific failure mode; however, a review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the product mp1000 china in the past 12 months.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Search Alerts/Recalls
|