|
A3b: gender: n/a.D6a: implanted date: device was not implanted.D6b: explanted date: device was not explanted.E2: health professional: requested, unknown.E3: occupation: requested, unknown.The actual sample was not available.Instead, a reference photo was received.The sheath for the sg3 needle is activated, but the needle is bent out of it.Retention samples were visually inspected and found to be free of any damage, bent needles, tilted cannula, or irregularities on the sheath and collar.There was no issued nonconformity report related to human error during lot production.The needles were manufactured at needle assembly machines 1 and 8.Machine 8 has an inspection system that detects tilted and bent cannulas.Dummy checking is conducted for every type of change, end of the lot, and machine adjustment/troubleshooting/replacement of parts to ensure the accuracy of the inspection system.During the collar assembly process, the hub fit is loosened from the protector before placing the needle into the collar.Prior to proceeding to the next process, the hub, cannula, and collar are pressed into the protector wherein a photoelectric sensor detects if the height of the protector is correct.Our safety needle features a hinged safety sheath attached to the needle hub.The safety sheath contains two locking mechanisms, the sheath tooth-cannula, and sheath wings-collar which are simultaneously activated when manually pressed over the needle immediately after use and just before disposal to minimize the possibility of sharps injury.At the cannula station, an inspector ensures proper cannula alignment to prevent bent needles.Our product has an allowable tilting of the needle of not more than 2°.Tilted cannulas are one of the check items in the hourly in-process inspection during needle assembly.The collar and sheath undergo an initial product inspection check (ipic) before mass production, during raw material change, mold maintenance, plant shutdown, cavity closure, and mold/product type change.Any molding defects that may affect product function are part of the inspection item.We also have a visual in-process and product confirmation inspection to ensure that the collar and sheath are in good condition.Visual in-process inspections are conducted during the manufacturing process to detect abnormalities in the product that may cause issues during sheath activation.The critical locking part of the sg3 product is checked for defects such as short shot, sheath collar fitting, and over or under insertion of the collar into the hub.We have functional tests conducted to confirm that our products are free of defects that would lead to difficulty or failure of device activation.We perform safety sheath activation during sg assembly inspection and outgoing inspection of finished products where the cannula should be captured under the sheath's tooth.During the component fit test, we also check that the sheath lugs are securely attached to the collar ear.Before shipment, qc conducts outgoing inspections to check the overall condition of the products.The product's instructions for use (ifu) provides detailed instructions for using the sg3 safety needle device, including warnings and precautions.Our cannula is a supplied raw material that complies with iso 9626, particularly on stiffness and breakage.It undergoes incoming inspection which includes visual inspection, dimensional inspection, and verification of quality certificate.The collar and sheath are manufactured in-house with validated tpc molding machines.There were no nonconformance reports issued for the supplied molded part lots.From the two fiscal years to the present, we received a total of nineteen (19) complaints for the specific item code.The cause of these complaints was not identified as being related to our product or the manufacturing process.The retention samples were injected into the dummy arm's veins and a rubber cork to simulate intravenous and intramuscular injections prior to safety activation.The safety sheath was successfully activated on the samples; the needle did not bend, and no difficulties were encountered.In the next simulation, the needle was forcefully injected into the cork, causing it to slightly bend.Despite being bent, the sheath-tooth was able to completely capture the cannula, and the collar wing was fully locked on the sheath, indicating that the activation was successful.The root cause of the problem could not be identified based on our investigation of all the available information regarding this issue.A review of the product's lot history file revealed no related issues that would cause the needle to bend during sheath activation.We have routine inspections to check the condition of the products.The components that are critical to the product's safety activation are checked to ensure that no defects occur that could result in sheath activation issues.We recommend following the instructions for use (ifu) for proper use of the sg3 needle, which include cautions, precautions, and warnings to avoid problems during sheath activation.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.3003902955.
|
|
The user facility reported that the needles bend after administration, resulting in two needle stick accidents.These accidents occurred post-treatment due to the bending of the needles.The impact on the user is currently unknown.Additional information was received on 20 dec 2024: the occurrence date of both needle stick injuries is on (b)(6) when we were notified about the incident.For each incident, we do not have information about whether the patient who received the injection was infected with hiv, hepatitis b, or another infectious disease.Consequently, it is not possible for us to determine if there is a risk that the user got infected after the needle stick, as this involves private data that we cannot access.
|