| Model Number |
6300 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One device was returned for analysis.Visual inspection found a malfunctioned downstream occlusion sensor.Functional and visual tests were performed and the reported issue was duplicated.The cause of the reported issue was due to the downstream occlusion sensor.As a result, the downstream occlusion sensor was replaced.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
|
| |
|
Event or Problem Description
|
|
It was reported that the pump was broken.During physical inspection, it was found that the downstream occlusion sensor was faulty.There was unknown patient involvement, no patient injury was reported.
|
| |
|