|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the objective lens cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the objective lens.In addition, our technician confirmed that the light guide cable buckled, the insertion flexible tube crushed, and the operation channel (primary) resistance; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|