EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-47-3009 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Osteolysis (2377)
|
| Date of Event |
08/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: 02-010-03-0330 - logic cr femoral cem, right, sz 3, (b)(6); 02-012-45-3030 - lgc tibial fit tray cem sz 3f / 3t, (b)(6); 200-02-35 - three peg patella 35mm, (b)(6).The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The reason for the revision reported cannot be confirmed from the information provided but may be the result of tibial loosening, prosthesis wear, and patellar loosening.Or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 44 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear and osteolysis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
|
| |
|
Search Alerts/Recalls
|
|
|