|
MEDOS INTERNATIONAL SÃ RL CH SET SCREW; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
|
Back to Search Results |
|
| Catalog Number |
102000000S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
01/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that almost all inner screws from a symphony construction detached from the head of the screw.The first surgery was on (b)(6) 2024 and in post op exam it was verified that that construction failed with the loose locking screws.Almost all pedicle screws remained on place.Non union was also noted.The patient underwent revision surgery, which went well.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6 the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found no damage or defect was found over the surface of the device.Additionally, a functional test was performed with the mating devices (b)(6) and no problem with assembly was found.Surgical technique symphony oct system was reviewed and the following relevant statements were found at page 13 and 14: note: the reducer (b)(6) helps ensure the rod is fully seated in the saddle before final tightening.Precaution: care should be taken to avoid using excessive force during rod reduction steps that may undermine bony purchase and result in screw pullout.Precautions: the screw inserters are not intended for use in final tightening.Damage to set screw driver feature may occur.Always fully reduce rod in all bone anchors prior to final tightening.Failure to use provided 3.0nm torque limiting handle when final tightening may lead to set screw back out.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was not confirmed as the observed condition of the set screw would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a manufacturing record evaluation was performed for the finished device.Product code: 102000000s.Lot number: 408868.The product was released on: 23 sep 2024.Manufacturing site: jabil le locle.Expiry date: 31 jul 2029.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|