| Catalog Number |
157001080 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hypersensitivity/Allergic reaction (1907); Pain (1994); Osteolysis (2377); Metal Related Pathology (4530)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the catalog/model number was not provided, the (01)gtin is not available.E3 initial reporter occupation: lawyer.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was implanted with a depuy asr hip implant in her right hip on (b)(6) 2007 and explanted on (b)(6) 2018 due to suffering from pain, injuries by excessive levels of chromium and cobalt.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown; therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received states that after review, the patient was revised due to failed right total hip arthroplasty w/ asr recall.Patient was experiencing pain and further exam found a pseudotumor.During the procedure there was a massive amount of osteolytic type of debris in the acetabulum, lysis found behind the acetabulum.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.According to the information received,¿ litigation records received ad 8 jan 2025.After review of litigation records, the patient was implanted with a depuy asr hip implant in her right hip on (b)(6) 2007 and explanted on (b)(6) 2018, due to suffering from pain, injuries by excessive levels of chromium and cobalt and economic loss.Doi: (b)(6) 2007, doe: (b)(6) 2018, right hip¿.Product description: - summit por taper sz2 std off.Product code: - 157001080.Lot no: - a66bn1000.Quantity of manufactured: - (b)(4).Date of manufacturing: - dec-2006.Expiry date: - nov-2016.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description: - summit por taper sz2 std off.Product code: - 157001080.Lot no: - a66bn1000.Quantity of manufactured: - (b)(4).Date of manufacturing: - dec-2006.Expiry date: - nov-2016.Device history review: a manufacturing record evaluation was performed for the finished device, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added : b6.If information that was not available for the initial report is obtained, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|