AOMORI OLYMPUS CO., LTD. SINGLE USE BALLOON DILATOR V (WITH KNIFE) Back to Search Results
Model Number BD-VC431Q-1240-25
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: 1.The cutting wire where the wire coating was torn came into contact with the distal end of the endoscope when the forceps elevator was raised.2.Under circumstances described above 1, an electric conduction was activated.This caused the cutting wire to become hot instantly at the contact point.As a result, the cutting wire broke.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that during a therapeutic est procedure, it was noticed that the est knife was broken before energizing.There was no harm or injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE BALLOON DILATOR V (WITH KNIFE)
Common Device Name
BALLOON DILATOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key21183580
Report Number9614641-2025-00070
Device Sequence Number8722116
Product Code KNS
UDI-Device Identifier04953170414107
UDI-Public04953170414107
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBD-VC431Q-1240-25
Device Lot Number45K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2024
Initial Report FDA Received Date01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-