C.R. BARD, INC. (BASD) -3006260740 DOT PROVPPSOLO 4FDL MBP; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
 
Event or Problem Description
It was reported that the needle from a double picc line kit safety mechanism malfunctioned.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a failed safety mechanism is confirmed; however, the exact cause is unknown.One photograph of a secureloc introducer needle was returned for evaluation.An initial visual observation of the photograph showed the introducer needle with the safety mechanism fully detached.Evidence of use was observed on the sample.No damage could be seen on the sample in the returned photograph.There were no distinguishing features in the returned photograph of the device which could aid in identification of a specific root cause.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported that the needle from a double picc line kit safety mechanism malfunctioned.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT PROVPPSOLO 4FDL MBP
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key21184023
Report Number3006260740-2024-08530
Device Sequence Number12919195
Product Code LJS
UDI-Device Identifier00801741155109
UDI-Public(01)00801741155109(11)241018(17)251031(10)REJW2729
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K180548
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberN/A
Device Catalogue NumberS1294108D4
Device Lot NumberREJW2729
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/27/2024
Supplement Date Received by Manufacturer03/14/2025
Initial Report FDA Received Date01/17/2025
Supplement Report FDA Received Date03/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-