• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM, ACCESSORY, EMBEDDED LASER; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM, ACCESSORY, EMBEDDED LASER; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751450
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
The government agency reported that during cataract surgery output energy from an ophthalmic system was too low while removing of the vitreous body from the patient left eye.The patient impact was not reported.Additional information has been requested but none received till date.
 
Additional Manufacturer Narrative
Additional information is provided in sections h.6 and h.11.The service history was reviewed for the system.No service record relevant to the reported event was found.All manufacturer surgical systems are verified to meet specifications after installation and customer initiated servicing in accordance with the applicable procedures.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM, ACCESSORY, EMBEDDED LASER
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21190129
Report Number2028159-2025-00072
Device Sequence Number11594666
Product Code HQC
UDI-Device Identifier00380657514502
UDI-Public(01)00380657514502(17)270712(11)170712(10)NA
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101285
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065751450
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/29/2024
Supplement Date Received by Manufacturer01/30/2025
Initial Report FDA Received Date01/20/2025
Supplement Report FDA Received Date02/11/2025
Date Device Manufactured07/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-