• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC EAKIN; PROTECTOR, OSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC EAKIN; PROTECTOR, OSTOMY Back to Search Results
Model Number 839002
Medical Device Problem Codes Melted (1385); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device 9 of 13.The end user made no mention of having any type of blockage.Hence, a1409 has not been selected in this emdr.Based on the available information, this event is deemed to be a reportable malfunction.Other company's product name: hollister 2 pcs convex.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Third party manufacturing site: 9681410.
 
Event or Problem Description
The end user mentioned that she had been using company's stoma seal since 2016.She usually gets seven days wear time.However, with the last two boxes, she had noticed that the ring was melted and got over the stoma within one day of use.At that point, she had to pick off the wax that was covering the opening in the stoma (os).Furthermore, the end user stated that when the wax was covering the opening in the stoma (os) she got liquid stool, and once the wax was removed from the opening in the stoma (os), the stool was darker in color when it came out.Since she had manually removed the wax, the stool was then back to its normal color and was toothpaste in consistency and that it felt well.She was now pushing on the ring with the appliance on, to ensure it was not covering the opening in the stoma (os).Her stoma was protruding well, and she was using another company's two-piece convex appliance with the company's ring.A personal support worker (psw) came weekly to assist with her changes.No photo is available at this time.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.Type of investigation: 4109 - 3331, investigation findings: 213, investigation conclusions: 4315.Supplier investigation, received, and the investigation found no faults in the production records and did not identify new defects.No samples were returned for testing, and no abnormalities were recorded in manufacturing paperwork.Previous complaints for this lot number reported similar issues with rings melting or being thinner and harder.The clinical team noted that a stoma appliance should not affect stool consistency and suggested other potential causes, such as digestive issues.The complaint did not meet regulatory reporting criteria, and corrective or preventive actions were deemed unnecessary.However, 13 adverse event reports were submitted to the food and drug administration (fda).This investigation will be closed.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).Fda registration number, reporting site: 1049092, third party manufacturing site: 9681410.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EAKIN
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC
7815 national service road
suite 600
greensboro NC 27409
MDR Report Key21190541
Report Number1049092-2025-00020
Device Sequence Number12912374
Product Code FON
UDI-Device Identifier00768455108732
UDI-Public00768455108732
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/30/2024
Report Date (Section F) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number839002
Device Lot Number109357C814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/30/2024
Date Report to Manufacturer01/10/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/20/2025
Supplement Report FDA Received Date06/11/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-