• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD MONOSOF; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD MONOSOF; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Model Number N-2730-K
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: n-2730-k, n-2730-k monosof, 10-0 blk 20cm se1406 (lot#: d4f2307y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a surgical procedure, when it was time to sew, the thread broke because it was too thin.Several attempts were made with other threads from the same lot, all of which also broke.A new suture was used to resolve the issue.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture was broken.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a surgical procedure, when it was time to sew, the thread broke because it was too thin.Several attempts were made with other threads from different batch, all of which also broke.A new suture was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: b5, d9, d10, g3, h3 correction: h6 d10 concomitant product: n-2730-k n-2730-k monosof* 10-0 blk 20cm se1406 - (lot#d4f2307y); n-2730-k n-2730-k monosof* 10-0 blk 20cm se1406 - (lot#d4e4101y); n-2730-k n-2730-k monosof* 10-0 blk 20cm se1406 - (lot#d4g3809y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOSOF
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21191247
Report Number9612501-2025-00175
Device Sequence Number14297418
Product Code GAR
UDI-Device Identifier10884521076266
UDI-Public10884521076266
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K900531
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN-2730-K
Device Catalogue NumberN-2730-K
Device Lot NumberD4F2307Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2024
Supplement Date Received by Manufacturer03/20/2025
05/14/2025
Initial Report FDA Received Date01/20/2025
Supplement Report FDA Received Date03/24/2025
05/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.; SEE NOTE ON H11.
Patient SexUnknown
-
-