| Model Number |
GIA6025S |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: "medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the firing knob of the instrument retracted and advanced properly without difficulty.The knife cut completely and cleanly and the staple line was properly formed.It was reported that the device component was disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to improper loading of the cartridge.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: for properly loading the sulu into the instrument, which includes holding the sulu by the proximal end tabs and inserting into the c artridge fork at a 30 to 45 degree angle from the distal end down until the unit snaps in place.The shipping wedge is removed after the sulu is fully loaded.The sulu will "float" up and down in its final position.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the blade of the cartridge popped up the moment the handle was installed and opened after the staple was mounted.The cartridge was loaded successfully but then popped off when the handle was attempted to be put together.Another cartridge was used but the same issue occurred.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d10, e1 (first name and last name), g3, h6 d10 concomitant product: unknown gia, unknown gia product, (lot # unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: unknown gia, unknown gia product (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the blade of the cartridge popped up the moment the handle was installed and opened after the staple was mounted.
|
| |
|