(b)(4).Date sent 1/21/2025 d4 batch #256c46 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned reload.Visual analysis of the returned sample determined that one ecr45g reload was received partially fired 1/4 with an open package.The returned reload was reset and loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples meet the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 256c46, and no non-conformances were identified.The lot#314c52 history records were reviewed and certified by external manufacturing that the manufacturing criteria were met prior to the release of the equipment.The certificate records are accessible through external manufacturing.
|