• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE PS RP INSRT SZ 3 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE PS RP INSRT SZ 3 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151650308
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 01/08/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient has an infection.Implants were removed and replaced with implant spacer.Doi: (b)(6) 2024.Dor: (b)(6) 2025.Affected side: left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code: 151650308, lot -3915541 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code: 151650308, lot -3915541 and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS RP INSRT SZ 3 8MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21199720
Report Number1818910-2025-01242
Device Sequence Number7578571
Product Code NJL
UDI-Device Identifier10603295052173
UDI-Public(01)10603295052173(17)300831(10)3915541
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151650308
Device Lot Number3915541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/08/2025
Supplement Date Received by Manufacturer02/11/2025
Initial Report FDA Received Date01/21/2025
Supplement Report FDA Received Date02/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-