EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-47-5011 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330)
|
| Date of Event |
01/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 8 years and 6 months post the initial right total knee arthroplasty, the patient reported back to the surgeon's office with complaints of stiffness, pain, and dissatisfaction with their right total knee.The patient was revised and underwent a tibial poly implant swap and patella implant swap.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.Photos provided show wear of the polyethylene insert.The product is not returning as it was discarded by the hospital.
|
| |
|
Additional Manufacturer Narrative
|
|
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.D10: (b)(6), 02-010-03-0350 - logic cr femoral cem, right, sz 5, (b)(6), 02-012-45-5050 - lgc tibial fit tray cem sz 5f / 5t, (b)(6), 200-02-32 - three peg patella 32mm.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4.This follow-up report is being submitted to relay corrected information.The following sections were corrected: d10.D10: 4206683 02-010-03-0350 - logic cr femoral cem, right, sz 5; 4161487 02-012-45-5050 - lgc tibial fit tray cem sz 5f / 5t.
|
| |
|
Event or Problem Description
|
|
Photos provided show wear of the polyethylene insert and patella.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h3, h6.H3: the revision reported was likely the result of tibial insert and patellar prosthesis wear.Possible causes for polyethylene wear include malalignment between the femoral component and the tibial insert, high contract stress during knee flexion, third body wear, patient-related conditions, inclusion of the polyethylene in the packaging recall with the insert being packaged for over five years prior to implantation, or any combination of these possibilities.
|
| |
|