• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES SAPIEN XT TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES SAPIEN XT TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9300TFX23A
Medical Device Problem Codes Degraded (1153); Perivalvular Leak (1457); Device Stenosis (4066)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 01/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by a field clinical specialist (fcs), approximately 9 years and 4 months post-implant of a 23 mm sapien xt valve in the aortic position, the valve failed due to stenosis.A valve in valve was successfully performed.
 
Additional Manufacturer Narrative
Per the instruction for use (ifu), valve stenosis is a potential risk associated with bioprosthetic heart valves.Per the valve academic research consortium (varc), valve stenosis can result from a number of factors, including pannus, calcification, support structure deformation (out-of-round configuration), trauma (cardia-pulmonary resuscitation, blunt chest trauma), endocarditis, prosthetic valve thrombosis, and native leaflet prolapse impeding prosthetic leaflet motion.Valve stenosis may result in symptoms such as shortness of breath and decreased exercise tolerance, which may be accompanied by an increased gradient across the valve.Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.Investigation results suggest the cause of the event is unknown due to the limited information available.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.
 
Additional Manufacturer Narrative
After reviewing the medical records provided, the patient had developed severe stenosis and severe transvalvular regurgitation of the prosthetic valve including pvl.This was confirmed by the intraoperative tee and the patient underwent a successful viviv procedure with a 23mm s3u valve.Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with the use of bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.Per the instructions for use (ifu), paravalvular leak (pvl) is a known potential adverse event associated with bioprosthetic heart valves and the transcatheter valve replacement (tvr) procedure.Paravalvular leak refers to blood flowing through a channel between the structure of the implanted valve and the landing zone, as a result of a lack of appropriate sealing of the valve to the target site.Some pvl is not uncommon post-deployment.Many cases are mild to moderate, and either resolve over time or do not cause symptoms.Others may be more clinically significant and require intervention.The patient screening manual and the procedure didactic identify several procedural and anatomical factors which could contribute to pvl, including device malposition, inaccurate measurement of the landing zone, uneven distribution of calcium on the landing zone, bulky or severe calcification, an elliptical annulus shape, and valve under-sizing.The mechanism behind worsening or late pvl is not well understood but may be related to cardiac remodeling.Edwards extensively trains physicians before they are qualified to use the sapien transcatheter heart valves (all models).The device training manuals instructs the operator on proper imaging screening requirements, including the use of good quality echocardiography and/or computed tomography (ct) to appropriately measure the landing zone, assess the content and distribution of calcium, and other patient anatomical factors.Multiple imaging modalities should be considered during valve size selection.Contraindications, important considerations when assessing the valve, and choosing the proper transcatheter heart valve are also discussed.The device training manuals also instruct the operator on proper positioning and deployment of the valve, including patients screening and procedural considerations.Device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures are also included.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest the cause of the event is unknown due to the limited information available.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAPIEN XT TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key21200669
Report Number2015691-2025-00417
Device Sequence Number12872482
Product Code NPT
UDI-Device Identifier00690103192070
UDI-Public(01)00690103192070(17)161209
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/09/2016
Device Model Number9300TFX23A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2025
Supplement Date Received by Manufacturer03/11/2025
Initial Report FDA Received Date01/21/2025
Supplement Report FDA Received Date03/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-