• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SMARTSET HIGH PERFORMANCE MV MEDIUM VISCOSITY BONE CEMENT 40G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SMARTSET HIGH PERFORMANCE MV MEDIUM VISCOSITY BONE CEMENT 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3122040
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 09/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a 55-year-old female patient received a second left knee revision to treat persistent global instability.The patient had a previous revision of the tibial insert to treat instability and poly-wear on (b)(6) 2018 ((b)(4)) upon entering the joint, the surgeon noted the tibial tray was stable but mildly loosened at the smart-set cement to implant interface.The femoral component was stable but revised.There was no reported product problem with the revised unknown tibial insert.The patella was well-fixed and retained.The patient received a depuy attune stemmed revision construct secured with competitor cement.The procedure was completed without complications.All available information has been disclosed.No further information was provided.Doi: (b)(6) 2018.Dor (b)(6) 2018 (primary tibial insert captured on (b)(4)).Dor: (b)(6) 2023 (unk insert, femoral, tray, cement).Left knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET HIGH PERFORMANCE MV MEDIUM VISCOSITY BONE CEMENT 40G
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key21201124
Report Number1818910-2025-01249
Device Sequence Number8534637
Product Code LOD
UDI-Device Identifier10603295168379
UDI-Public(01)10603295168379(17)190430(10)8532927
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K081155
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Catalogue Number3122040
Device Lot Number8532927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/08/2025
Supplement Date Received by Manufacturer01/23/2025
Initial Report FDA Received Date01/21/2025
Supplement Report FDA Received Date01/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-