Catalog Number 314-13-13
Medical Device Problem Codes Fracture (1260); Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Osteolysis (2377)
Date of Event 12/31/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3: customer has indicated that the product is in process of being returned for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that this patient's right shoulder was revised.Patient had pain that lead to x-rays.Observation showed migration on the pe and the peripheral pegs.Only the central peg was still in place, it was initially cemented.The patient was converted to a reverse with a baseplate+10mm and bone graft.Patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: h3: the revision reported was likely the result of the glenoid loosening and fracture of the center cage leading to migration and wear of the polyethylene body.However, the reason for the loosening and fracture could not be confirmed, and contributions from incomplete/improper seating of the device and patient-related considerations to the event could not be assessed from the information provided.The observed damage and wear on the articular surface of the glenoid component may have been due to contact with the humeral bone and/or damage following migration of the glenoid component.The observed wear on the humeral head was likely due to unintended contact with the retained center cage following fracture of the center polyethylene peg and subsequent migration of the glenoid body.The following sections were corrected: h6: medical device problem code, type of investigation, investigation findings, investigation conclusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID MEDIUM, BETA
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key21203395
Report Number1038671-2025-00329
Device Sequence Number8071542
Product Code KWS
UDI-Device Identifier10885862172693
UDI-Public10885862172693
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314-13-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2024
Supplement Date Received by Manufacturer04/16/2025
Initial Report FDA Received Date01/21/2025
Supplement Report FDA Received Date05/06/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight65 KG
-
-