• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + DUET ASYMMETRICAL NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + DUET ASYMMETRICAL NASAL CANNULA Back to Search Results
Model Number OPT964
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.Product background: the opt964 optiflow + duet asymmetrical nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.
 
Event or Problem Description
A healthcare facility in the united kingdom reported via a fisher & paykel healthcare (f&p) field representative that an opt964 optiflow + duet asymmetrical nasal cannula was found damaged and leaking during patient use.The subject device was then replaced.There were no reported patient consequences.
 
Event or Problem Description
A healthcare facility in the united kingdom reported via a fisher and paykel healthcare (f&p) field representative that an opt964 optiflow + duet asymmetrical nasal cannula was found damaged before patient use.The healthcare facility stated that the cannula tubing had split.There was no patient involvement as the issue was found whilst the device was not in use on a patient.
 
Additional Manufacturer Narrative
(b)(4).Corrected data: a1, b5, d8, g3, g6, h2, h6.The healthcare facility reported that the subject opt964 optiflow + duet asymmetrical nasal cannula was discarded.Product background: the opt964 optiflow + duet asymmetrical nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.Method: the complaint opt964 optiflow + duet asymmetrical nasal cannula was not returned to f&p for evaluation.F&p's investigation is based on the information provided by the healthcare facility, previous investigations of similar complaints, and f&p's knowledge of the product.Results: the healthcare facility has stated that the opt964 optiflow + duet asymmetrical nasal cannula was damaged.The healthcare facility stated that the cannula tubing had split.There was no patient involvement reported by the healthcare facility.Conclusion: f&p's investigation was unable to determine the exact cause of the reported damage.Based on f&p's knowledge of the product, the reported damage is likely to have been caused by the tubing being subjected to excessive force.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The subject opt964 optiflow + duet asymmetrical nasal cannula would have met the required specifications.The user instructions which accompany the opt964 optiflow + duet asymmetrical nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state: "do not crush or stretch tube, to prevent loss of therapy." "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + DUET ASYMMETRICAL NASAL CANNULA
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key21203579
Report Number9611451-2025-00064
Device Sequence Number8156977
Product Code BTT
UDI-Device Identifier09420012469564
UDI-Public(01)09420012469564(10)2103137946(11)240516
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K231956
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT964
Device Catalogue NumberOPT964
Device Lot Number2103137946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2024
Supplement Date Received by Manufacturer02/14/2025
Initial Report FDA Received Date01/21/2025
Supplement Report FDA Received Date03/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-